Microscopy
Studying this course? Keep your progress in one place.
Key facts
- Magnification = image size ÷ actual size; convert to the same units first.
- Resolution is the minimum distance between two points seen as separate; more magnification does not improve it.
- Light microscopes resolve about 0.2 µm; electron microscopes use electrons with a shorter wavelength for far higher resolution.
- TEM shows internal structure in 2D; SEM shows the surface in 3D. Electron microscopes need dead specimens in a vacuum.
- A graticule must be calibrated with a stage micrometer for each objective lens.
Magnification and resolution
Magnification makes the image larger; resolution is how much detail you can see.
Magnification is how many times larger the image is than the object. Rearranged, actual size = image size ÷ magnification. Convert to the same units first (1 mm = 1000 µm; 1 µm = 1000 nm).
Resolution is the minimum distance between two points that can be seen as separate. Enlarging beyond the resolution limit gives only a bigger, blurred image: empty magnification.
- 1
Measure the image
e.g. 45 mm
- 2
Convert units
45 mm × 1000 = 45 000 µm
- 3
Divide by magnification
45 000 ÷ ×300 = 150 µm
- 4
Actual size
150 µm (0.15 mm)
- Magnification= image size ÷ actual size
- Actual size= image size ÷ magnification
Worked example
A cell image is 24 mm long at ×3000. What is the actual length of the cell?
- 1
Convert the image size: 24 mm = 24 000 µm.
- 2
Actual size = image size ÷ magnification.
- 3
24 000 ÷ 3000 = 8.
What is resolution?
Light microscopes
A light microscope resolves about 0.2 µm, limited by the wavelength of light.
A light microscope uses visible light and glass lenses. Resolution is about 0.2 µm (200 nm), limited by the wavelength of light, and useful magnification is up to about ×1500.
Advantages: specimens can be living, preparation is quick and cheap, and colour stains can be used.
It cannot resolve small structures such as ribosomes.
Advantages
- Living specimens
- Quick and cheap
- Colour
Disadvantages
- Low resolution (0.2 µm)
- Cannot see ribosomes
Why does a light microscope have a limited resolution?
Electron microscopes
Electrons have a much shorter wavelength than light, so electron microscopes give far higher resolution.
An electron microscope uses a beam of electrons with a much shorter wavelength than light. Electromagnets focus the beam, and the specimen must be in a vacuum, so it must be dead.
A TEM passes electrons through a thin section, showing internal ultrastructure in 2D. An SEM scans the surface, giving a 3D image at lower resolution.
Disadvantages: expensive, complex preparation, black and white images and artefacts.
- 1
Electron gun
fires a beam in a vacuum
- 2
Electromagnets
focus the beam
- 3
Thin section
electrons pass through or are scattered
- 4
Screen
the image forms on a fluorescent screen
| Light microscope | TEM | SEM | |
|---|---|---|---|
| Radiation | Visible light | Electron beam | Electron beam |
| Resolution | About 0.2 µm | About 0.1 nm | Lower than TEM |
| Specimen | Can be living | Dead, in a vacuum | Dead, in a vacuum |
| Image | Colour, 2D | Black and white, 2D, internal | Black and white, 3D, surface |
Light microscope
- Radiation:
- Visible light
- Resolution:
- About 0.2 µm
- Specimen:
- Can be living
- Image:
- Colour, 2D
TEM
- Radiation:
- Electron beam
- Resolution:
- About 0.1 nm
- Specimen:
- Dead, in a vacuum
- Image:
- Black and white, 2D, internal
SEM
- Radiation:
- Electron beam
- Resolution:
- Lower than TEM
- Specimen:
- Dead, in a vacuum
- Image:
- Black and white, 3D, surface
Which microscope shows the surface of a specimen in 3D?
The importance of staining
Most structures are almost transparent, so stains increase contrast.
Most cell structures have little contrast. A stain binds to particular structures so they absorb more light and can be told apart.
For electron microscopy, heavy metals such as lead or uranium compounds scatter electrons and give contrast.
- 1
Unstained
structures are almost transparent, little contrast
- 2
Stain binds
to particular structures
- 3
Absorbs more
stained structures absorb light (or scatter electrons)
- 4
Clear image
structures can be told apart
Methylene blue or acetic orcein
- Nucleus and DNA
Iodine in potassium iodide
- Starch (blue-black)
Eosin
- Cytoplasm
Lead or uranium compounds
- Electron microscopy contrast
Why is iodine in potassium iodide used on plant tissue?
Core Practical 5: drawing and graticule
Calibrate an eyepiece graticule against a stage micrometer, then convert divisions into real lengths.
Temporary mount: place thin tissue in a drop of water or stain and lower the coverslip at an angle to avoid air bubbles.
Drawing: focus on low power, then high. Use a sharp pencil, single continuous lines, no shading, ruled label lines, a title and the magnification.
A graticule has arbitrary units. Calibrate it with a stage micrometer (divisions of 10 µm) for each objective lens.
- 1
Line up the scales
graticule against the stage micrometer
- 2
Calibrate
work out the length of one graticule division
- 3
Replace the micrometer
put the specimen slide on the stage
- 4
Measure
count graticule divisions across the cell
- 5
Convert
divisions × length of one division
Worked example
80 graticule divisions line up with 400 µm on the stage micrometer. A cell is 14 graticule divisions long. What is its length?
- 1
One division = 400 µm ÷ 80 = 5 µm.
- 2
Cell length = 14 × 5 µm.
Why is the graticule recalibrated for each objective lens?
Try an exam question
Explain why an electron microscope can show more detail than a light microscope, and give one disadvantage of using an electron microscope.
[3 marks]
- [1]Electrons have a much shorter wavelength than light.
- [1]So the electron microscope has a higher resolution.
- [1]Disadvantage: specimens must be dead (in a vacuum), or images are black and white, or artefacts may form.
That's the notes covered.
Carry on to the next subtopic.
Try a question
What is the actual diameter of the cell?
1 mark, marked against the mark scheme
What is the actual diameter of the cell?
1 mark, marked against the mark scheme