Microscopy

Microscopy

Spec 3.7 · Edexcel International A Level Biology: Revision notes

Studying this course? Keep your progress in one place.

Key facts

  • Magnification = image size ÷ actual size; convert to the same units first.
  • Resolution is the minimum distance between two points seen as separate; more magnification does not improve it.
  • Light microscopes resolve about 0.2 µm; electron microscopes use electrons with a shorter wavelength for far higher resolution.
  • TEM shows internal structure in 2D; SEM shows the surface in 3D. Electron microscopes need dead specimens in a vacuum.
  • A graticule must be calibrated with a stage micrometer for each objective lens.

Magnification and resolution

Magnification makes the image larger; resolution is how much detail you can see.

Magnification is how many times larger the image is than the object. Rearranged, actual size = image size ÷ magnification. Convert to the same units first (1 mm = 1000 µm; 1 µm = 1000 nm).

Resolution is the minimum distance between two points that can be seen as separate. Enlarging beyond the resolution limit gives only a bigger, blurred image: empty magnification.

  1. 1

    Measure the image

    e.g. 45 mm

  2. 2

    Convert units

    45 mm × 1000 = 45 000 µm

  3. 3

    Divide by magnification

    45 000 ÷ ×300 = 150 µm

  4. 4

    Actual size

    150 µm (0.15 mm)

Finding actual size: worked numbers.
  • Magnification= image size ÷ actual size
  • Actual size= image size ÷ magnification

Worked example

A cell image is 24 mm long at ×3000. What is the actual length of the cell?

What is resolution?

Light microscopes

A light microscope resolves about 0.2 µm, limited by the wavelength of light.

A light microscope uses visible light and glass lenses. Resolution is about 0.2 µm (200 nm), limited by the wavelength of light, and useful magnification is up to about ×1500.

Advantages: specimens can be living, preparation is quick and cheap, and colour stains can be used.

It cannot resolve small structures such as ribosomes.

10020030040050060070080050100150200250300350400xy400 nm light: 200 nmresolution ≈ λ ÷ 2
Resolution is about half the wavelength used. Visible light of about 400 nm resolves about 200 nm (0.2 µm).

Advantages

  • Living specimens
  • Quick and cheap
  • Colour

Disadvantages

  • Low resolution (0.2 µm)
  • Cannot see ribosomes

Why does a light microscope have a limited resolution?

Electron microscopes

Electrons have a much shorter wavelength than light, so electron microscopes give far higher resolution.

An electron microscope uses a beam of electrons with a much shorter wavelength than light. Electromagnets focus the beam, and the specimen must be in a vacuum, so it must be dead.

A TEM passes electrons through a thin section, showing internal ultrastructure in 2D. An SEM scans the surface, giving a 3D image at lower resolution.

Disadvantages: expensive, complex preparation, black and white images and artefacts.

  1. 1

    Electron gun

    fires a beam in a vacuum

  2. 2

    Electromagnets

    focus the beam

  3. 3

    Thin section

    electrons pass through or are scattered

  4. 4

    Screen

    the image forms on a fluorescent screen

Path of electrons through a TEM.

Light microscope

Radiation:
Visible light
Resolution:
About 0.2 µm
Specimen:
Can be living
Image:
Colour, 2D

TEM

Radiation:
Electron beam
Resolution:
About 0.1 nm
Specimen:
Dead, in a vacuum
Image:
Black and white, 2D, internal

SEM

Radiation:
Electron beam
Resolution:
Lower than TEM
Specimen:
Dead, in a vacuum
Image:
Black and white, 3D, surface

Which microscope shows the surface of a specimen in 3D?

The importance of staining

Most structures are almost transparent, so stains increase contrast.

Most cell structures have little contrast. A stain binds to particular structures so they absorb more light and can be told apart.

For electron microscopy, heavy metals such as lead or uranium compounds scatter electrons and give contrast.

  1. 1

    Unstained

    structures are almost transparent, little contrast

  2. 2

    Stain binds

    to particular structures

  3. 3

    Absorbs more

    stained structures absorb light (or scatter electrons)

  4. 4

    Clear image

    structures can be told apart

Why staining helps.

Methylene blue or acetic orcein

  • Nucleus and DNA

Iodine in potassium iodide

  • Starch (blue-black)

Eosin

  • Cytoplasm

Lead or uranium compounds

  • Electron microscopy contrast

Why is iodine in potassium iodide used on plant tissue?

Core Practical 5: drawing and graticule

Calibrate an eyepiece graticule against a stage micrometer, then convert divisions into real lengths.

Temporary mount: place thin tissue in a drop of water or stain and lower the coverslip at an angle to avoid air bubbles.

Drawing: focus on low power, then high. Use a sharp pencil, single continuous lines, no shading, ruled label lines, a title and the magnification.

A graticule has arbitrary units. Calibrate it with a stage micrometer (divisions of 10 µm) for each objective lens.

  1. 1

    Line up the scales

    graticule against the stage micrometer

  2. 2

    Calibrate

    work out the length of one graticule division

  3. 3

    Replace the micrometer

    put the specimen slide on the stage

  4. 4

    Measure

    count graticule divisions across the cell

  5. 5

    Convert

    divisions × length of one division

Measuring a cell with a graticule

Worked example

80 graticule divisions line up with 400 µm on the stage micrometer. A cell is 14 graticule divisions long. What is its length?

Why is the graticule recalibrated for each objective lens?

Try an exam question

Explain why an electron microscope can show more detail than a light microscope, and give one disadvantage of using an electron microscope.

[3 marks]

That's the notes covered.

Carry on to the next subtopic.

Try a question

What is the actual diameter of the cell?

1 mark, marked against the mark scheme

Revision notes